Disposable biosensor based on ionic liquid, carbon nanofiber and poly (glutamic acid) for tyramine determination
Küçük Resim Yok
Tarih
2024
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Academic Press Inc Elsevier Science
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
In this study, an electrochemical biosensor based on carbon nanofibers (CNF), ionic liquid 1-butyl-3-methylimidazolium hexafluorophosphate (IL), poly(glutamic acid) (PGA) and tyrosinase (Tyr) modified screen printed carbon electrode (SPE) was constructed for tyramine determination. Optimum experimental parameters such as CNF and IL amount, polymerization conditions of glutamic acid, enzyme loading, pH of test solution and operating potential were explored. The construction steps of the Tyr/PGA/CNF-IL/SPE were pursued by scanning electron microscopy and cyclic voltammetry. The Tyr/PGA/CNF-IL/SPE biosensor exhibited linear response to tyramine in the range of 2.0 x 10-7 - 4.8 x 10-5 M with a low detection limit of 9.1 x 10-8 M and sensitivity of 302.6 mu A mM-1. The other advantages of Tyr/PGA/CNF-IL/SPE include its high reproducibility, good stability and anti-interference ability. The presented biosensor was also applied for tyramine determination in malt drink and pickle juice samples and mean analytical recoveries of spiked tyramine were calculated as 100.6% and 100.4% respectively.
Açıklama
Anahtar Kelimeler
Amperometry, Biosensor, Tyramine, Carbon nanofiber, Ionic liquid, Polyglutamic acid
Kaynak
Analytical Biochemistry
WoS Q Değeri
Q2
Scopus Q Değeri
Q3
Cilt
684